4. Unscrew the five M3×6 screws that fasten the AC input receptacle assembly and then the screw that fastens the
grounding cable. Unplug the AC input receptacle and cable from the board to remove the AC input receptacle
assembly.
5. Unscrew the three M3×6 screws to remove the AC/DC power board.
7.3.9 Removing the Power Management Board
Unscrew the seven M3×6 screws inside the monitor, disconnect all the cables, and then remove the power
management board.
AC input receptacle and cable