12 Measuring Procedure
40 / 92 Instruction Manual Myopia Master® (G/68100/EN Rev04 0820)
Pachymeter values (3)
P. Apex: Corneal thickness at the apex
P. Min: Corneal thickness at the thinnest point
Q-value:
If the field has a white background (9-8) - the measuring results are
good.
If the field has a yellow background (7-6) - the measuring results are
critical; repeat the measurement, if necessary.
If the field has a red background (
6) - repeat the measurement.
IOP Chg.: Correction of the intraocular pressure dependent on the
corneal thickness in the apex (sec. 17, page 62).
Correction is done based on the selected correction table. The type
of correction table can be selected in the settings.
Scheimpflug (4)
The Scheimpflug image shows the cornea between the red doted lines.
Keratometer Values (5)
Rh/Rv: Horizontal/vertical radius of curvature in the center
blue: flat meridian
red: steep meridian
Pupil: Size of pupil
Astig: Astigmatism of cornea in the center
WTW: (white-to-white) Cornea diameter or iris diameter.
Q-value:
If the field has a white background (9-8) - the measuring results are
good.
If the field has a yellow background (7-6) - the measuring results are
critical; repeat the measurement, if necessary.
If the field has a red background (
6) - repeat the measurement.
Refraction Values (6)
The sphere, cylinder, axis position and quality values are displayed in this
field.
The refraction values are measured three times. The mean value is
displayed in the fourth line.
Q-value:
If the field has a white background (9-8) - the measuring results are good.
If the field has a yellow background (7-6) - the measuring results are
critical; repeat the measurement, if necessary.
If the field has a red background (
6) - repeat the measurement.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the OCULUS Myopia Master Advanced and is the answer not in the manual?

OCULUS Myopia Master Advanced Specifications

General IconGeneral
BrandOCULUS
ModelMyopia Master Advanced
CategoryMedical Equipment
LanguageEnglish

Related product manuals