24 Annex
Instruction Manual Myopia Master® (G/68100/EN Rev04 0820) 85 / 92
Electromagentic immunity, IEC 60601-1-2: 2015, based on table 4
Immunity test IEC 60601
test level
Compliance level Electromagnetic environment -
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV
± 15 kV
Floors should be made of wood
or concrete or covered with
ceramic tiles. If the floor is co-
vered with synthetic material,
the relative humidity must be at
least 30%.
Power frequency (50/60Hz)
magnetic field
IEC 61000-4-8
30 A/m
50 Hz or 60 Hz
30 A/m
50 Hz or 60 Hz
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital envi-
ronment.
Electromagentic immunity, IEC 60601-1-2: 2015, based on table 5, 8
Electrical Fast transient/bursts
IEC 61000-4-4
± 2 kV for power
supply lines
100 kHz repe-
tition frequency
± 1 kV for input/
output lines
± 2 kV
-----------------
± 1 kV
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
± 1 kV
kine(s) to line(s)
± 2 kV
line(s) to earth
± 1 kV
differential mode
± 2 kV
line(s) to earth
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short interruptions,
and voltage variations on power
supply input lines
IEC 61000-4-11
0% U
τ
; 1/2 period
at 0, 45, 90, 135,
180, 225, 270 and
315 degree
0% U
τ
; 1 period
and
70% U
τ
; 25/30
periods
Single-phase: at
0 degree
0% U
τ
; 250/300
periods
0% U
τ
; 1/2 period
at 0, 45, 90, 135,
180, 225, 270 and
315 degree
0% U
τ
; 1 period
and
70% U
τ
; 25/30
periods
Single-phase: at
0 degree
0% U
τ
; 250/300
periods
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the Myopia Master® requires
continued operation during
power mains interruptions, it is
recommended that the Myopia
Master® be powered from an
uninterruptible power supply or
battery.
Note: U
τ
is the a.c. mains voltage prior to application of the test level.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the OCULUS Myopia Master Advanced and is the answer not in the manual?

OCULUS Myopia Master Advanced Specifications

General IconGeneral
BrandOCULUS
ModelMyopia Master Advanced
CategoryMedical Equipment
LanguageEnglish

Related product manuals