24 Annex
Instruction Manual Myopia Master® (G/68100/EN Rev04 0820) 87 / 92
Recommended separation distances between portable and mobile
RF communications equipment and the Myopia Master®,
IEC 60601-1-2:2007, table 6
The Myopia Master® in intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the Myopia Master® can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters)
and the Myopia Master® as recommended below, according to the maximum output power of the communica-
tions equipment.
Rated maximum output pow-
er of transmitter
W
Separation distance according to frequency of transmitter m
150 KHz to 80 Mhz
d= 1.2 P
80 MHz to 800 MHz
d= 1.2 P
800 MHz to 2.5 GHz
d= 2.3 P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.80 3.80 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d
in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is
the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the OCULUS Myopia Master Advanced and is the answer not in the manual?

OCULUS Myopia Master Advanced Specifications

General IconGeneral
BrandOCULUS
ModelMyopia Master Advanced
CategoryMedical Equipment
LanguageEnglish

Related product manuals