CH10/CH20 C. DISASSEMBLY AND ASSEMBLY PROCEDURES
C-6
2. Revolving Nosepiece
No. Parts name Screw Grease Adhesive Remarks
SCREW
NOSEPIECE
SPRING
Absolute optical axis: within 0.3
mm
Click removing force: 6.4 - 8.8N
{650 - 900g}
Optical axis adjustment of
revolving nosepiece (See Repair
Procedure, D-1.)
ƒ
SHAFT ANU5X6SA (*1)
B3SO (*2) 10 pcs.
OT2008 OT1028 Apply grease on the balls
(B3SO) fitting surface.
Apply adhesive on the screw
thread.
NOSEPIECE B3SO (*3) 66 pcs. OT2008 Apply grease on the balls
(B3SO) fitting groove.
RING
KNOB OT2008 Apply grease on the screw
thread.