34
Recommended separation distance between portable and mobile RF communications
equipment and the OMRON BP7100
OMRON BP7100 is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of this OMRON BP7100 can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the OMRON BP7100 as recommended below,
according to the maximum output power of the communications equipment.
Output Power of Transmitter in
Watt
Separation distance according to frequency of transmitter
in meter
150 kHz to 80 MHz
Not Applicable.
80 MHz to 800 MHz
d = 1.2
800 MHz to 2.5GHz
d = 2.3
0.01
Not Applicable.
0.12 0.23
0.1 0.38 0.73
11.22.3
10 3.8 7.3
100 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
Note: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.