1RQ,QYDVLYH3UHVVXUH0HDVXUHPHQW3ULQFLSOHV
Oscillometric method
The beat in the pulsation generated by the contraction of the heart is captured
as the pressure inside the cuff to measure the blood pressure. If the cuff
wrapped around the upper arm is pressurized sufficiently, the blood flow
stops, but the beat of the pulsation is present and the pressure inside the cuff
receives this and oscillates. Next, as the pressure inside the cuff gradually
decreases, the oscillation of the pressure within the cuff gradually increases
and reaches a peak. As the pressure within the cuff decreases further, the
oscillation decreases from its peak.
The pressure within the cuff and the relationship with the increase and
decrease of the oscillation within the cuff in this series of processes are
stored into memory, calculations are carried out, and the blood pressure
value is determined.
The oscillometric method does not determine the blood pressure value
instantly like a microphone type automatic blood pressure gauge with the
auscultation method, but rather determines it from the series of change
curves as explained above. Therefore, it is not easily affected by external
noise, an electric scalpel or other electro surgical instruments.

Other manuals for Omron HBP-1120

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omron HBP-1120 and is the answer not in the manual?

Omron HBP-1120 Specifications

General IconGeneral
BrandOmron
ModelHBP-1120
CategoryBlood Pressure Monitor
LanguageEnglish

Related product manuals