Check Standards
Kemio is supplied with check standards (KEMC10) to confirm
the instrument is working correctly.
Validating Kemio Using Check Standards
Start a check standard test by
A) pressing CONFIGURATION then CHECK STANDARDS
or
B) pressing PERFORM TEST or open the arm, and selecting
CHECK STANDARDS
(Check standards must be enabled as a test type.)
3
The check takes 10
seconds. If the instrument
is functioning correctly, a
PASS will display on screen
and the corresponding line
will turn green.
If Kemio displays a FAIL,
repeat the test with the same
standard. If the problem
persists, contact your Palintest
representative.
4
Press NEXT or open the arm to
test the next standard. Repeat
with each standard in this way.
If all standards pass, Kemio will
display a summary report and
return to the home screen.
Results are stored in the data
log for traceability.
Open the arm, and Insert the
first of four check standards
from the KEMC10 pack.
2
1
1
PASS
STANDARD S1
STANDARD S2
STANDARD S3
STANDARD S4
STANDARD S1
SUCCESSFUL
PASS
STANDARD S1
STANDARD S2
STANDARD S3
STANDARD S4
STANDARD S1
SUCCESSFUL