Separation distance according to frequency of transmitter m
Rated maximum
output power of
transmitter W
150 kHz to 80
MHz outside ISM
bandsd=1,17 √P
80 MHz to 800
MHzd=1,17 √P
800 MHz to 2,5 GHz
d=2,33 √P
0,01 0,12 0,12 0,23
0,1 0,38 0,38 0,73
1 1,2 1,2 2,3
10 3,8 3,8 7,3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in metres (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2:These guidelines may not apply in all situations. Electromagnetic propagation is aected by
absorption and reection from structures, objects and people.
Usage and storage conditions
Keep the breast pump out of direct sunlight as prolonged exposure may cause discoloration.
Store the breast pump and its accessories in a safe, clean and dry place.
If the appliance has been stored in a hot or cold environment, place it in the usage environment to let
it reach a temperature within the usage conditions ( 41°F / 5°C to 104°F / 40°C ) before you use it.
Usage conditions
Temperature 41°F (+5°C) to 104°F (+40°C)
Relative humidity 15% to 93% (non-condensing)
Atmospheric pressure 700 and 1060 Pa air pressure
Storage conditions
Temperature -13°F (-25°C) to 158°F (70°C)
Relative humidity 15% to 93% (non-condensing)