SimplyGo User Manual
EMC Information
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
This device is intended for use in the electromagnetic environment specied below� The user of this device should assure that it
is used in such an environment�
Emissions Test Compliance Electromagnetic Environment - Guidance
RF Emissions
CISPR 11
Group 1 The Device uses RF energy only for its internal
function� Therefore, its RF emissions are very
low and are not likely to cause any interference
in nearby electronic equipment�
RF Emissions
CISPR 11
Class B The Device is suitable for use in all
establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network
that supplies buildings used for domestic
purposes�
Harmonic Emissions
IEC 61000-3-2
Class A
Voltage Fluctuations/Flicker Emissions
IEC 61000-3-3
Complies
Emission of Radio Frequency Energy
RTCA/DO-160G Section 21
Category M The Device is suitable for use onboard
commercial airplanes inside passenger cabin�
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
This device is intended for use in the electromagnetic environment specied below� The user of this device should assure that it
is used in such an environment�
Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment - Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±8 kV contact
±2 kV, ±4 kV, ±8 kV, &
±15 kV Air
±8 kV contact
±2 kV, ±4 kV, ±8 kV, &
±15kV Air
Floors should be wood, concrete or ceramic tile�
If oors are covered with synthetic material, the
relative humidity should be at least 30%�
Electrical Fast
Transient/Burst
IEC 61000-4-4
±2 kV for Power Supply
Lines
±1 kV for Input/Output
Lines
±2 kV for Power Supply
Lines
NA- Device does not have
user I/O lines that are
longer than 3m in length�
Mains power quality should be that of a typical
home or hospital environment�
Surge
IEC 61000-4-5
±1 kV Line to Line
±2 kV Line to Ground
±1 kV Line to Line
NA- The device is a Class
II device and does not
connect to earth ground�
Mains power quality should be that of a typical
home or hospital environment�
Voltage Dips, Short
Interruptions and
Voltage Variations on
Power Supply Input
Lines
IEC 61000-4-11
0% U
T
0�5 cycle,
45deg inc
0% U
T
1 cycle
70% U
T
25 Cycles
(30 cycles if US)
0% U
T
5 sec
0% U
T
0�5 cycle,
45 deg inc
0% U
T
1 cycle
70% U
T
25 Cycles
(30 cycles if US)
0% U
T
5 sec
Mains power quality should be that of a typical
home or hospital environment� If the user of the
Device required continued operation during
power mains interruptions, it is recommended
that the Device be powered from an
uninterruptible power supply or battery�
Power Frequency
(50/60 Hz) Magnetic
Field
IEC 61000-4-8
30 A/m 30 A/m Power frequency magnetic elds should be at
levels characteristic of a typical location in a
typical home or hospital environment�
Note: U
T
is the A�C� mains voltage prior to application of the test level�