28
SimplyGo User Manual
Connecting the Optional Humidier
Warnings
• Onlyusethehumidierconnectortubingsuppliedwiththe
SimplyGoHumidierConnectingTubekit(REF1101172)orthe
SimplyGoHumidierPouchandTubekit(REF1101602).
• Donotoverllhumidier.
• Donotreversetheoxygeninputandoutputconnections.Waterfrom
the humidier bottle will travel through the cannula back to the patient�
• DonotusethehumidierwhilethedeviceisinPulsemode.
Note: Only connect the humidier if prescribed� Only use the humidier if the
deviceisinContinuousFlowmode.
To connect the humidier:
1� Attach the humidier accessory pouch to the mobile cart (see the
instructions included with the pouch)�
2� Remove the cap from the humidier bottle�
3. Fillthehumidierwithdistilledwatertothelevelindicatedbythe
manufacturer�
4� Replace the cap and make sure it is tightened securely�
5� Attach the humidier bottle adapter tube to the humidier bottle
by turning the wing nut on the bottle counterclockwise until it is
fastened securely�
6. Placethehumidierbottleassemblyintheaccessorypouch(REF
1083699)� Attach the accessory bag to the pull cart�
7� Rotate the bottle so the humidier adapter hose is away from the
body of the portable oxygen concentrator�
8� Attach the oxygen tubing from the humidier bottle assembly to the
oxygen outlet connector on the portable oxygen concentrator�
9� Attach the cannula tubing to the humidier bottle outlet�
10� After assembly is complete, make sure the oxygen is owing through
the cannula�
Notes
• AbubblehumidiermaybeusedwiththeSimplyGo.
• AHumidierAccessoryPouch(REF1083699)maybepurchasedto
enable use of a bubble humidier�