EasyManuals Logo
Home>pico Technology>Test Equipment>PicoScope 2205A

pico Technology PicoScope 2205A Programmer's Guide

Default Icon
55 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #40 background imageLoading...
Page #40 background image
PicoScope 2000 Series Programmer's Guide 43
Copyright © 2006–2016 Pico Technology Ltd. All rights reserved. ps2000pg.en r11
5.25
ps2000_set_ets
int32_t ps2000_set_ets
(
int16_t handle,
int16_t mode,
int16_t ets_cycles,
int16_t ets_interleave
)
This function is used to enable or disable ETS mode and to set the ETS parameters.
Applicability
Not PicoScope 2202
Arguments
handle: the handle of the required oscilloscope
mode:
PS2000_ETS_OFF (0) - disables ETS
PS2000_ETS_FAST (1) - enables ETS and provides
ets_cycles cycles of data, which may contain data from
previously returned cycles
PS2000_ETS_SLOW (2) - enables ETS and provides fresh data
every ets_cycles cycles. PS2000_ETS_SLOW takes longer to
provide each data set, but the data sets are more stable and
unique
ets_cycles: the number of cycles to store. The computer can then
select ets_interleave cycles to give the most uniform spread of
samples. ets_cycles should be between two and five times the
value of ets_interleave.
ets_interleave: the number of ETS interleaves to use. If the
sample time is 20 ns and the interleave 10, the approximate time
per sample will be 2 ns.
Returns
The effective sample time in picoseconds, if ETS is enabled
0: if ETS is disabled or one of the parameters is out of range

Other manuals for pico Technology PicoScope 2205A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the pico Technology PicoScope 2205A and is the answer not in the manual?

pico Technology PicoScope 2205A Specifications

General IconGeneral
Brandpico Technology
ModelPicoScope 2205A
CategoryTest Equipment
LanguageEnglish

Related product manuals