Simplified instruction for the Graphic Chart recording time
8.6 Print timer/ Communication Channel NO. (Option)
Press the SET key to enter the password “9”, Press the Set key again to enter
print-time/Communication channel NO. set function. Press the Increase or Decrease
key to modify the print-timer/Communication channel NO. Press the Set key to con-
firm/store the new setting.
Simplified instructions for the print timer (Option)
8.7 Motor fan speed adjustment
Forced air convection may destroy the protective layer from moist air that is generated
during incubation over the samples. This would result in dehydration of the culture.
Provided the chamber is fully loaded and forced air circulation is required, this unit is
installed with main ventilating fan speed adjusting option, to provide further flexibility for
various applications.
The speed can be adjusted by 3 levels, High (Factory Default), Medium and Low.
To set the ventilator speed:
Press the SET key and enter the password “13”. Press the SET key again until the LCD
display shows the fan speed menu, press the Increase or Decrease key to select the
desired speed level. To confirm the new settings press the SET key again.
Special Attention:
This function is subject to your choosen option of either printer or
RS485 option.
This will be used for
-printing interval timer setting, if printer is installed,
-communication channel NO. setting, if RS485 interface is installed.
Graphic chart
Interval time
Print Interval Time/
Communication Channel
NO.