INCU-C Series V2018.1.1
21
Simplified instructions for adjusting fan speed
9.Operation and Switch Off
9.1 When all the above settings are done, press the “R/H/S” button and the equipment
will start running according to the program settings.
9.2 If pressing the “R/H/S” button while the unit is running, the timer will pause until the
“R/H/S” button is pressed again.
9.3 If the “R/H/S” button is held for 3 seconds, the remaining run time will be cleared to
zero. Press the “R/H/S” button once more, and the unit will start again to count down
from the preset operation time.
9.4 While the unit is in use, the current remaining operating time cannot be changed. If
however changed at this time, it is invalid with the current operation. Only when the
current operation time is elapsed- or stopped and activated again according to the
above instructions, the new changed value will be effective.
9.5 The incubator can be turned off with the main Power switch on the right side of the
unit to end operation completely.
Special Attention:
Fan Speed Adjustment may slightly affect the temperature uniformity
across the working chamber.
The fan speed can be further adjusted upon request, please contact
us or your local agent for more information.