INCU-C Series V2018.1.1
23
11.6 Keep pressing "SET" button, skip the " High Temperature Point Correction (100.0)"
setting, save and exit.
11.7 The displayed temperature should have changed due to the calibration. Waiting for
another one hour to let the temperature stabilize again at 8°C , and check the ther-
mometer value again, and calculate the new difference.
11.8 if necessary, perform the calibration again until the actual display value equals to the
calibrated thermometer value. Thus the Step 1. "Low Temperature Point Correction
(0.0)" is completed.
High Temperature Point Correction (100.0)
11.9 If the incubator is to be used for more than one temperature setting, and "High
Temperature Point Correction (100.0)" needs to be performed as well.
11.10 Change the set point to a higher temperature point as required, for example
60.0°C and let the incubator run for at least 1 hour– until the temperature is constant,
and let the temperature inside of the chamber is uniformed
11.11 Change the enter the code "47" again and press "SET" skip the "0.0", and enter
the "100.0" page.
11.12 Repeat the same procedure, like the "Low Temperature Point Correction", to
correct the high temperature point, until the display value equals to the actual
thermometer value. Then the "Step 2 High Temperature Point Correction (100.0)" is
completed.