INCU-C Series V2018.1.1
4
1. The unit must be placed horizontally on a solid, flat floor or table.
2. The incubator needs even heat lost on all surfaces in order to maintain small internal
temperature variations. As a result, a minimum of 20 cm must be allowed between the rear and
sides of the incubator and any walls, partitions or obstructions to facilitate adequate convection
of air around the unit.
3. Do not locate the unit exposed to direct sunlight or near heating /cooling ducts.
4. The unit must be kept away from electromagnetic interference sources.
5. Culture placed inside the incubator should be placed to the extent of which the air flow inside
the incubator is not affected to maintain the temperature uniformity in the working chamber.
6. Make sure the shelves inside the chamber are placed horizontally.
7. Slam the door(s) will probably leads to damage of the equipment.
8. When in operation, do not open the door (too much or too long) as this might affect the
temperature inside.
9. The incubator must be kept away from volatile, flammable, explosive liquids or gases.
10. Please keep the chamber clean. Regular cleaning is required.
11. After transport of the Incubator – DO NOT USE IT for at least 24 Hours, unless you are a
100% sure that the unit has been moved in an upright position only.
12. If the incubator is used continuously at low temperatures, con-
densate can occur inside of the incubator. You might need to wipe this
out, or-if possible-heat up the incubator to evaporate the condensation.
There is however on the right side (low) a drain that might be emptied.
(see the picture).
13. To extend the compressors life and maintain an excellent perfor-
mance of your refrigeration system, the condenser of the unit should be
cleaned every month.
Attention!
Apart from the above warnings and instructions, there are several
other special notes that need attention.
Please read them. Any neglect might cause serious problems,
damage, or injury.
Attention!
Instructions for optimal performance