3. Click Diagnostics & Calibration.
4. In the Diagnostics and Calibration screen, click Calibration.
All calibration routines are selected by default and cannot be deselected.
5. Click Start.
The Start button changes to Abort once the calibration has started. Clicking Abort stops the calibration process and
restores all values to their previous state. All calibration items should indicate Pass. If not, contact your local Tektronix
service personnel.
Install the instrument
AWG5200 Series Installation and Safety Instructions 11