Commands Listed in Alphabetical Order
Syntax
ETHERnet:HTTP
Port <QSt ring>
ETHERnet:HTTPPort?
Arguments
<QString> is an integer port number, enclosed in quotes.
NOTE. Consider the following if you are using the e*Scopeā„¢ control software. If
you don't enter a port address in the URL, then the
ETHERnet:HTTPPort value
must be set to "80", which is the default port for HTTP protocol. If you use a
URL with a port address (for example: http://DPO2004-04WKL4:1234), the port
number is specified by the number after the colon. Set the
ETHERnet:HTTPPort
valuetothissamenumber.
Examples
ETHERNET:HTTPPORT ā€œ80ā€ sets the HTTP port value to 80.
ETHERnet:IPADDress
Sets or returns the IP address assigned to the oscilloscope.
Group
Ethernet
Syntax
ETHERnet:IPADDress <QString>
ETHERnet:IPADDress?
Arguments
<QString> is a standard IP address value, enclosed in quotes.
Examples
ETHERNET:IPADDRESS ā€œ123.121.13.214ā€ sets the oscilloscope's IP address.
ETHERnet:NAME
Sets or returns the network name assigned to the oscilloscope.
Group
Ethernet
Syntax
ETHERnet:NAME <QString>
ETHERnet:NAME?
Arguments
<QString> is the network name assigned to the oscilloscope, enclosed in quotes.
2-150 MSO2000B, DPO2000B, M SO2000 and DPO2000 Series Oscilloscopes Programmer Manual

Other manuals for Tektronix DPO2012B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO2012B and is the answer not in the manual?

Tektronix DPO2012B Specifications

General IconGeneral
BrandTektronix
ModelDPO2012B
CategoryTest Equipment
LanguageEnglish

Related product manuals