Syntax
MEASUrement:REFLevel:PERC ent:MID2 <NR3>
MEASUrement:REFLevel:PERC ent:MID2?
Related Commands
MEASUrement:REFLevel:METHod
Arguments
<NR3> is the mid reference level, ranging from 0 to 100%. The default mid
reference level is 50%.
Examples
MEASUREMENT:REFLEVEL:PERC ENT:MID2 40 s ets the mid2 reference level
to 40% of HIGH.
MEASUREMENT:REFLEVEL:PERC ENT:MID2? might return
:MEASUREMENT:REFLEVEL:PER CENT:MID2 45 indicating that the
percentage mid2 reference level is set to 45% of HIGH.
MEASUrement:REFLevel:PERCent:MID<x>
Sets or returns the mid reference level for channel <x>, where x is the
measurement channel.
Group
Measurement
Syntax
MEASUrement:REFLevel:PERC ent:MID<x> <NR3>
MEASUrement:REFLevel:PERC ent:MID<x>?
Arguments
<NR3> is the mid reference level in percent.
MEASUr
ement:SNAPShot (No Query Form)
Displays the measurement snapshot list on the oscilloscope screen. The list
contains the immediate values for all available measurements of the active signal.
Group
Measurement
Syntax
MEASUrement:SNAPShot
2-214 MSO2000B, DPO2000B, M SO2000 and DPO2000 Series Oscilloscopes Programmer Manual