Commands Listed in Alphabetical Order
value sets the p
ower-on status clear flag to true, enabling the power-on status clear
preventing any SRQ assertion after power on.
Examples
*PSC 0 sets the power-on status clear fl ag to false.
*PSC? might return 1 to indicate that the power-on status clear flag is set to true.
*PUD
Sets or returns a string of Protected User Data. This d ata is protected by the
PASSWord command. You can modify it only by first entering the correct
password. This password is not necess ary to query the data.
Group
Status and Error
Syntax
*PUD {< Block>|<QString> }
*PUD?
Related Commands
PASSWord
Arguments
<Block> is a block containing up to 300 ASCII characters.
<QString> is a string containing up to 300 ASCII characters.
Examples
*PUD #229This oscilloscope belongs to me stores the string "This
oscilloscope belongs to me" in the user protected data area.
*PUD? might return #221PROPERTY OF COMPAN Y X
*RCL (No Query Form)
This command restores the state of the oscilloscope from a copy of the settings
stored in memory (The settings are stored using the *SAV command).
Group
Save and Recall
Syntax
*RCL <N R1>
Related Commands
FACtory, *LRN?, RECAll:SETUp, *RST, *SAV, SAVe:SETUp
MSO2000B, DPO2000B, MSO2000 and DPO2000 Series Oscilloscopes Programmer Manual 2-227