Commands Listed in Alphabetical Order
Syntax
SELect:D<x> {<
NR1>|OFF|ON}
SELect:D<x>?
Arguments
ON turns on the display of the specified waveform. This waveform also becomes
the selected waveform.
OFF turns o ff the display of the specified waveform.
<NR1> = 0 turns off the display of the specified waveform; any other value turns
on the display of the specified waveform.
Examples
SELECT:D2 ON turns the digital channel 2 waveform display on, and selects
digital channel 2.
SELECT:D2? might return :SELECT:D2 1 indicating that digital channel 2 is
being displayed.
SELect:MATH[1]
Turns on and off the display of the math waveform. The query returns whether
the math waveform is on or off but does not indicate whether it is the selected
waveform.
Group
Vertical
Syntax
SELect:MATH[1] {ON|OFF|<NR1>}
SELect:MATH[1]?
Arguments
ON turns on the display of the specified waveform. This waveform also becomes
the selected waveform.
OFF turns o ff the display of the specified waveform.
<NR1> = 0 turns off the display of the specified waveform; any other value turns
on the display of the specified waveform.
Examples
SELECT:MATH ON turns the math waveform display on, and selects it.
SELECT:MATH? might return : SELECT:MATH 1 indicating that the math
waveform is being displayed.
MSO2000B, DPO2000B, MSO2000 and DPO2000 Series Oscilloscopes Programmer Manual 2-287