EasyManuals Logo
Home>Tektronix>Test Equipment>DPO2012B

Tektronix DPO2012B Programmer's Manual

Tektronix DPO2012B
441 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #396 background imageLoading...
Page #396 background image
Status and Events
Table 3-1: SESR
Bit Functions
Bit F unction
7 (MSB) PON Power On. Show
s that the oscilloscope was powered on.
On completion, the diagnostic s elf tests also set this bit.
6
URQ
User Request
.
Indicates that an application event has
occurred. *See note.
5
CME
Command Error. Shows that an error occurred while
the oscilloscope was parsing a command or query.
4 EXE
Execution Error. Shows that an error executing a
command or
query.
3DDE
Device Err
or. Shows that a device error occurred.
2
QYE
Query Error. E ither an attempt was made to read the
Output Qu
eue when no data was present or pending, or that
data in the Output Queue was lost.
1
RQC
Request C
ontrol.
This is not used.
0 (LSB) OPC Operation Complete. Shows that the operation
is complete. This bit is set when all pending operations
comple
te following an *O PC command.
The Status Byte Register (SBR). Records whether output is available in the Output
Queue, whether the oscilloscope requests service, and whether the SESR has
recorded any events.
Use a Serial Poll or the *STB? query to read the contents of the SBR. The bits in
the SBR are set and cleared depending on the contents of the SESR, the Event
Status Enable Register (ESER), and the Output Queue. When you use a Serial
Poll to obtain the SBR, bit 6 is the RQS bit. When you use the *STB? query to
obtain the SBR, bit 6 is the MSS bit. Reading th
e SBR does not clear the bits.
Figure 3-2: The Status Byte Register (SBR)
Table 3-2: SBR Bit Functions
Bit Function
7(MSB)
————
Not used.
6
RQS
Request Service. Obtained from a serial poll. Shows
that the o scilloscope requests service from the GPIB
controller.
6
MSS
Master Status Summary. Obtained from *STB?
query. Summarizes the ESB and MAV bits in the SBR.
5
ESB
Event Status Bit. Shows that status is enabled and
present in the SESR.
3-2 MSO2000B, DPO2000B, MSO2000 and DPO2000 Series Oscilloscopes Programmer Manual

Other manuals for Tektronix DPO2012B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO2012B and is the answer not in the manual?

Tektronix DPO2012B Specifications

General IconGeneral
BrandTektronix
ModelDPO2012B
CategoryTest Equipment
LanguageEnglish

Related product manuals