Appendix: Warranted Specications
DC Balance 0.2 div with the input DC-50 coupled and 50 terminate d
0.25 div at 2 mV/div with the input DC-50 coupled and 50 terminated
0.5 div at 1 mV/div with the input DC-50 coupled and 50 terminated
0.2 div with the input DC-1 M coupled an d 50 terminated
0.3 div at 1 mV/div with the input DC-1 M coupled and 50 terminated
DC gain accuracy For 1 M path:
±1.5%, derated at 0.100%/°C above 30 °C
±3.0% Variable Gain, derated at 0.100%/°C
above 30 °C
For 50 path:
±1.5%, d erated at 0.050%/°C above 30 °C
±3.0% Variable Ga in, derated at 0.050%/°C
above 30 °C
Offset accuracy ±[0.005 × | offset position | + DC Balance]
NOTE. Both the position and constant offset term must be converted to volts by multiplying by the
appropriate volts/div term.
Long-term sample
rate and delay time
accuracy
±5 ppm over any 1 ms time interval.