Appendix: Warranted Specications
Appendix: Warranted Specications
The limits stated b elow are for ambient temperature of 30 °C and the bandwidth selection set to FULL.
Reduce the upper bandwidth frequency by 1% for each °C abo ve 30 °C.
Instrument
5 m V/div to 1 V/div 2 mV/div to 4.98 mV/div 1 mV/div to
1.99 mV/div
DPO/MSO4104 DC to 1 GHz DC to 350 MHz DC to 200 MHz
DPO/MSO4054 DC to 500 MHz DC to 350 MHz DC to 200 MHz
Instrument
2mV/divto1V/div 1mV/divto1.99V/div
DPO/MSO4034 DC to 350 MHz DC to 200 MHz
Analog bandwidth,
50
DPO/MSO4032 DC to 350 MHz DC to 200 MHz
Input impedance, DC
coupled
1M ±1%inparallelwith13p2pF
50 ±1%
DPO/MSO4101: VSWR 1.5:1 f rom DC to 1 GHz, typical
DPO/MSO4054: VSW R 1.5:1 from DC to 500 MHz, typical
DPO/MSO4034, DPO/MSO4032: VSWR 1.5:1 from DC to 350 MHz, typical
MSO4000 and DPO4000 Series Oscilloscopes User Manual 319