Commands Listed in Alphabetical Order
*IDN? (Query O
nly)
Returns the power supply identification code in IEEE 488.2 notation.
Group
Status
Syntax
*IDN?
Returns
A string that includes <manufacturer>, <model>, <serial number>, and
<firmware_version> as defined in the following table.
<manufacturer> <model> <serial number> <firmware_version>
keithley 22XXX XXXXXX X. XX-X. XX
Examples
*IDN?
might return the following response for a 2220-30-1:
KEITHLEY , 2220-30–1 , 000004 , 1.01–1.20
INSTrument:COMbine? (Query Only)
This command queries the instrument to det
ermine the combination state of
channels 1 and 2.
Group
Channel
Syntax
INSTrument:COMbine?
Related Commands
INSTrument:COMbine:OFF
Returns
Series for series combination.
Parallel for parallel combination.
NONE for combination off.
INSTrument:COMbine:OFF
This command is used to turn off series, parallel, or tracking mode, and return
channels 1 and 2 to independent operation.
Series 2200 Programmable Multichannel DC Power Supplies Programmer Manual 2-17