Commands Listed in Alphabetical Order
Syntax
[SOURce]:CHAN
nel:OUTPut:[STATe] {0|1|ON|O FF}
[SOURce]:CHANnel:OUTPut:[STATe]? ON
Related Commands
INSTrument:SELect
Arguments
0 or OFF indicates that the channel is off.
1 or ON indicates that the c hannel is on.
Examples
CHANNEL:OUTPUT OFF would turn whichever channel had been selected using
an
INSTRUMENT command to the off state.
CHAN:OUTP? might return 0 to indicate that the current channel is off.
[SOURce:]CURRent[:LEVel]:DOWN[:IMMediate][:AMPLitude] (No Query Form)
This command is used to decrease the current level by a step.
The stepping current can be set by the following command:
[SOURce:]CURRent[:LEVel][:IMMediate]:STEP[:INCRement]
Group
Source
Syntax
[SOURce:]CURRent[:LEVel]:DOWN[:IMMediate][:AMPLitude]
Related Commands
[SOURce:]CURRent[:LEVel][:IMMediate]:
STEP[:INCRement]
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]
This command is used to set or query the current value of the power supply in
units of A or mA.
Group
Source
Syntax
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]
{<Current>|MIN|MAX}
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]?
Related Commands
INSTrument:SELect
2-26 Series 2200 Programmable Multichannel DC Power Supplies Programmer Manual