Commands Listed in Alphabetical Order
[SOURce:]APP
ly (No Query Form)
This command is used to select channels and set voltage and current level using a
single command.
Group
Source
Syntax
[SOURce:]APPly
{CH1|CH2|CH3},[Voltage|Max|Min],[Current|Max|Min]
[SOURce:]APPly [CH1|CH2|CH 3]
Related Commands
This command can replace the following commands: INST CH1, VOLT 3V,
and CUR R 1A. (See example below.)
Arguments
CH1, CH 2, or CH3 are the three channels (two for two channel instrument
models).
Voltage.
Voltage is a flexible decimal number (NRf) that may be type NR1,
NR2 or NR3. It specifies the voltage setting, which can range from 0 to the
maximum nameplate voltage of the power supply.
MAX sets the voltage to th
e m aximum level (note that the maximum level may be
somewhat higher than the nameplate).
MIN sets the voltage to the minimum level (0 V).
Current.
Current is a flexible decimal number (NRf) that may be type NR1,
NR2 or NR3. It specifies a level between the minimum current and maximum
nameplate current level for the power supply.
MAX sets the current to the maximum level.
MIN sets the current to the minimum level (0 A).
Examples
[SOURCE:]APPLY CH 1,3V,1A would set channel 1 to 3 volts, 1 amps.
[SOURce]:CHANnel:OUTPut:[STATe]
This command is used to individually control the output state of the single,
currently-specified channel.
Group
Source
Series 2200 Programmable Multichannel DC Power Supplies Programmer Manual 2-25