Commands Listed in Alphabetical Order
Group
Source
Syntax
[SOURce:]VOLTage:LIMit:STATe {0|OFF|1|ON}
[SOURce:]VOLTage:LIMit:STATe?
Related Commands
INSTrument:SELect
[SOURce:]VOLTage:LIMit[:LEVel]
Arguments
0 or OFF turns off the maximum voltage limit.
1 or ON turns it on.
Examples
VOLTAGE:LIMIT:STATE ON
[SOURce:]VOLTage:TRIGgered[:IMMediate]
This command is used to set the voltage level for the trigger function. The
command queues the next setting for the currently selected channel. The units
are kV, V, mV or μV. When the instrument receives its next trigger, the c urrently
selected channel will be set to the specified value. So, 8000000 μVwillsetthe
voltage to 8 V on the front panel.
Group
Source
Syntax
[SOURce:]VOLTage:TRIGgered[:IMMediate] {<vol tage>|MIN|MAX}
[SOURce:]VOLTage:TRIGgered[:IMMediate]?
Related Commands
INSTrument:SELect
Arguments
<voltage> is a flexible decimal number (NRf) that may be type NR1, NR2 or
NR3. It specifies a level between the minimum voltage and maximum nameplate
voltage level for the power supply.
MIN sets the voltage to the minimum level (0 V).
MAX sets the voltage to the maximum level.
Examples
VOLTAGE:TRIGGERED 4.5V
2-36 Series 2200 Programmable Multichannel DC Power Supplies Programmer Manual