:DATA?
(:CALibration:PROTected:SENSe:DATA?)
(:CALibration:PROTected:SOURce:DATA?)
Purpose To request the calibration constants for the active range.
Format :cal:prot:sens:data?
:cal:prot:sour:data?
Response Four comma-separated ASCII floating-point constants.
Description The :CAL:PROT:SENS:DATA? and :CAL:PROT:SOUR:DATA? queries
request the calibration constants for the active range of the sense and source
functions respectively. The four returned constants are in ASCII floating-
point format delimited by commas.
Note To request the appropriate constants, (1) select the source or sense function,
(2
) cho
ose the desired range, and (3) send the desired :DATA? query.
Example :SENS:FUNC 'VOLT:DC' Select DC sense function.
:SENS:VOLT:DC:RANGE 0.2 Choose 200mV range.
:CAL:PROT:SENS:DATA? Request cal constants.
B-10 Command Reference
2410-902-01.book Page 10 Monday, November 7, 2005 2:49 PM

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2410 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2410 Specifications

General IconGeneral
BrandTektronix
ModelKeithley SourceMeter 2410
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals