EasyManuals Logo
Home>Tektronix>Measuring Instruments>Keithley SourceMeter 2410

Tektronix Keithley SourceMeter 2410 User Manual

Tektronix Keithley SourceMeter 2410
125 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #16 background imageLoading...
Page #16 background image
1-6 Performance Verification
Restoring factory defaults
Before performing the verification proced
ures, restore the instrument to its factory front panel
(bench) defaults as follows:
1. P
ress the MENU key. The instrument will display the following prompt:
MAIN MENU
SAVESETUP COMMUNICATION CAL
2. Select SAVESETUP, and then press ENTER. The unit displays:
SAVESETUP MENU
GLOBAL SOURCE-MEMORY
3. Select GLOBAL, and then press ENTER. The unit then displays:
GLOBAL SETUP MENU
SAVE RESTORE POWERON RESET
4. Select RESET, and then press ENTER. The unit displays:
RESET ORIGINAL DFLTS
BENCH GPIB
5. Select BENCH, and then press ENTER. The unit then displays:
RESETTING INSTRUMENT
ENTER to confirm; EXIT to abort
6. Press ENTER to restore bench defaults, and note the unit displays the following:
RESET COMPLETE
BENCH defaults are now restored
Press ENTER to continue
7. Press ENTER and then EXIT as necessary to return to normal display.
2410-902-01.book Page 6 Monday, November 7, 2005 2:49 PM

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2410 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2410 Specifications

General IconGeneral
BrandTektronix
ModelKeithley SourceMeter 2410
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals