4. Once the correct password is displayed, press the ENTER key. If the password was
correctly entered, the following message will be displayed:
CALIBRATION UNLOCKED
Calibration can now be executed.
5. Press EXIT to return to normal display. Calibration will be unlocked and will assume
the states summarized in Table 2-2. Attempts to change any of the settings listed below
with calibration unlocked will result in an error +510, "Not permitted with cal
unlocked."
NOTE With
calibratio
n unlocked, the sense function and range track the source function and
range. That is, when :SOUR:FUNC is set to VOLT, the :SENS:FUNC setting will be
'VOLT:DC'. When :SOUR:FUNC is set to CURR, the :SENS:FUNC setting will be
'CURR:DC'. A similiar command coupling exists for :SOUR:VOLT:RANG/
:SENS:VOLT:RANG/ and SOUR:CURR:RANG/:SENS:CURR:RANG.
Table 2-2
Calibration unlocked states
Mode State Equivalent remote command
Concurrent Functions
Sense Function
Sense Volts NPLC
Sense Volts Range
Sense Current NPLC
Sense Current Range
Filter Count
Filter Control
Filter Averaging
Source V Mode
Volts Autorange
Source I Mode
Current Autorange
Autozero
Trigger Arm Count
Trigger Arm Source
Trigger Count
Trigger Source
OFF
Source
1.0
Source V
1.0
Source I
10
REPEAT
ON
FIXED
OFF
FIXED
OFF
ON
1
Immediate
1
Immediate
:SENS:FUNC:CONC OFF
:SENS:FUNC <source_ function>
:SENS:VOLT:NPLC 1.0
:SENS:VOLT:RANG <source_V_range>
:SENS:CURR:NPLC 1.0
:SENS:CURR:RANG <source_ I_ range>
:SENS:AVER:COUN 10
:SENS:AVER:TCON REPeat
:SENS:AVER:STAT ON
:SOUR:VOLT:MODE FIXED
:VOLT:RANG:AUTO OFF
:SOUR:CURR:MODE FIXED
:SOUR:CURR:RANG:AUTO OFF
:SYST:AZERO ON
:ARM:COUNT 1
:ARM:SOUR IMMediate
:TRIG:COUNT 1
:TRIG:SOUR IMMediate
2-6 Calibration
2410-902-01.book Page 6 Monday, November 7, 2005 2:49 PM

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2410 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2410 Specifications

General IconGeneral
BrandTektronix
ModelKeithley SourceMeter 2410
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals