Remote calibration procedure
Step 1: Prepare the Model 2410 for calibration
1. Connect the Model 2410 to the
controller IEEE-488 interface or RS-232 port using a
shielded interface cable.
2. Turn on the Model 2410 and the test equipment, and allow them to warm up for at least
one hour before performing calibration.
3. If you are using the IEEE-488 interface, make sure the primary address of the Model
2410 is the same as the address specified in the program you will be using to send
commands. (Use the MENU key and the COMMUNICATION menu to access the
IEEE-488 address.)
Table 2
-7
Recommended :CALibration:PROTected:SOURce parameter ranges
Source
range
Firs
t parameter
(negative full scale)
Second parameter
(negative zero)
Third parameter
(positive full scale)
Fourth parameter
(positive zero)
0.2V
2V
20V
1000V
1µA
10µA
100µA
1mA
20mA
100mA
1A
-0.18 to -0.22
-1.8 to -2.2
-18 to -22
-800 to -1200
-0.9E-6 to -1.1E-6
-9E-6 to -11E-6
-90E-6 to -110E-6
-0.9E-3 to -1.1E-3
-18E-3 to -22E-3
-90E-3 to -110E-3
-0.9 to -1.1
-0.002 to +0.002
-0.02 to +0.02
-0.2 to +0.2
-2 to +2
-1E-8 to +1E-8
-1E-7 to +1E-7
-1E-6 to +1E-6
-1E-5 to +1E-5
-1E-4 to +1E-4
-1E-3 to +1E-3
-1E-2 to +1E-2
+0.18 to +0.22
+1.8 to +2.2
+18 to +22
+800 to +1200
+0.9E-6 to +1.1E-6
+9E-6 to +11E-6
+90E-6 to +110E-6
+0.9E-3 to +1.1E-3
+18E-3 to +22E-3
+90E-3 to +110E-3
+0.9 to +1.1
-0.002 to +0.002
-0.02 to +0.02
-0.2 to +0.2
-2 to +2
-1E-8 to +1E-8
-1E-7 to +1E-7
-1E-6 to +1E-6
-1E-5 to +1E-5
-1E-4 to +1E-4
-1E-3 to +1E-3
-1E-2 to +1E-2
Note: Parameter steps for each range may be performed in any order, but all four parameter steps for each range must be
completed. For optimum calibration, use parameters within recommended limits.
2-18 Calibration
2410-902-01.book Page 18 Monday, November 7, 2005 2:49 PM

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2410 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2410 Specifications

General IconGeneral
BrandTektronix
ModelKeithley SourceMeter 2410
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals