EasyManuals Logo
Home>Tektronix>Test Equipment>MDO3014

Tektronix MDO3014 User Manual

Tektronix MDO3014
243 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #218 background imageLoading...
Page #218 background image
Appendix A: Warr
anted Specications
Table 1 : RF Channel Characteristics (cont.)
Characteristic Description
Spurious Response (SFDR)
2
nd
harmonic distortion: >100 MHz, < –55 dBc
2nd harmonic distortion: > 100 MHz, < –60 dBc (typical)
With auto settings on and signals 10 dB below reference level
2
nd
harmonic distortion:9 kHz to 100 MHz, < –55 dBc
2
nd
harmonic distortion: 9 kHz to 100 MHz, < –60 dBc ( typical)
With auto settings on and signals 10 dB below reference level and reference level
–15 dBm
3
rd
order harmonic distortion: >100 MHz < –53 dBc
3
rd
order harmonic distortion: >100 MHz < –58 dBc (typical)
With auto settings on and signals 10 dB below reference level
3
rd
order harmonic distortion: 9 kHz to 100 MHz, < –55 dBc
3
rd
order harmonic distortion: 9 kHz to 100 MHz, < –60 dBc (typical)
With auto settings on and signals 10 dB below reference level and reference level
–15 dBm
2
nd
order intermodulation distortion: >100 Mhz, < –55 dBc
2
nd
order intermodulation distortion: >100 MHz, < –60 dBc (typical)
With auto settings on and signals 10 dB below reference level
2
nd
order intermodulation distortion: 9 k Hz to 15 MHz, < –55 dBc
2
nd
order intermodulation distortion: 9 kHz to 15 MHz, < –60 dBc (typical)
With auto settings on and signals 10 dB below reference level and reference level
–15 dBm
3
rd
order intermodulation distortion: >15 MHz, < –55 dBc
3
rd
order intermodulation distortion: >15 MHz < –60 dBc (typical)
With auto settings on and signals 10 dB below reference level
3
rd
order intermodulation distortion:9 kHz to 15 MHz, < –55 dBc
3
rd
order intermodulation distortion: 9 kHz to 15 MHz, < –60 dBc (typical)
With auto settings on and signals 10 dB below reference level and reference level
–15 dBm
–45 dBc (–50 dBc, typical) for side bands < 25 kHz offset from the carrier
Residual Spurious Response
< –78 dBm
< –67 dB m at 2.5 GHz
< –76 dBm at 1.25 GHz
–15 dBm reference level and RF input terminated with 50 .
Reference Frequency Error
(Cumulative)
Cumulative Error: ±10 x 10
-6
Includes allowances for Aging per Year, Reference Frequency Calibration Accuracy,
and Temperature Stability.
Valid over the recommended 1 year calibration interval, from –10 °C to +55 °C .
196 MDO3000 Series Oscilloscopes User Manual

Table of Contents

Other manuals for Tektronix MDO3014