EasyManuals Logo
Home>Tektronix>Test Equipment>MDO3014

Tektronix MDO3014 User Manual

Tektronix MDO3014
243 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #217 background imageLoading...
Page #217 background image
Appendix A: Warr
anted Specications
Delta Time
Measurement
Accuracy
The formula to calculate delta-time measurement accuracy (DTA) for a given instrument setting and input
signal is given below (assumes insignicant signal content above Nyquist)
SR
1
=SlewRate
(1
st
Edge) around the 1
st
point in the m easurement
SR
2
=SlewRate(2
nd
Edge) around the 2
nd
point in the m easurement
N = input-referred noise (volts
rms
, Refer to the Random Noise, Sample acquisition mode specication)
t
sr
=1/(Samp
le Rate)
TBA = timebase accuracy (Refer to the Long-term sample rate and delay time accuracy speci cation)
t
p
= delta-time measurement duration (sec)
RD = (R ecord
Length) / (Sample Rate)
Assumes that error due to aliasing is insignicant.
The term under the square-root sign is the stability, and is due to TIE (Time Interval Error). The errors due
to this term occur throughout a single-shot measurement. The second term is due to both the absolute
center-frequency accuracy and the center-frequency stability of the timebase and varies between multiple
single-shot measurements over the observation interval (the amount of time from the rst single-shot
measurement to the nal single-shot measurement).
Threshold
Accuracy, digital
input
+/- [100 mV + 3% of threshold setting after calibration]
Requires valid SPC.
Table 1: RF Channel Characteristics
Characteristic Description
Phase Noise
10 kHz: < –81 dBc/Hz; (–85 dBc/Hz, typical)
100 kHz: < –97 dBc/Hz; (–101 dBc/Hz, typical)
1 MHz: < –118 dBc/Hz; (–122 dB c/Hz, typical)
Phase Noise measured offset from 1 GHz CW signal.
Displayed Average Noise Level
(DANL)
9 kHz to 50 kHz: < –109 dBm/Hz; (< –113 dBm/Hz, typical)
50 kHz to 5 MHz: < –126 dBm/Hz; ( –130 dB m/Hz, typical)
5 MHz to 2 GHz : < –138 dBm/Hz; (< –142 dBm/Hz, typical)
2 GHz to 3 GHz: < –128 dBm/Hz; (< –132 dBm/Hz, typical)
Level Measurement Uncertainty
< ±1.2 dB, < ±0.6 dB (typical), 18 °C - 28 °C temperature range
< ±2.0 dB, –10 °C to +55 °C
MDO3000 Series Oscilloscopes User Manual 195

Table of Contents

Other manuals for Tektronix MDO3014