Clear and Sanitize Procedures
Memory Devices
The following tables list the volatile and nonvolatile memory devices in the
standard instrument and listed options. Detailed procedure s t o clear or sanitize
these devices, if any, are shown following each table.
Table 1: Volatile memory devices
Type and minimum
size Function
Type
of user
info
stored
Backed
up by
battery
Method
of
modi-
cation
Data input
method Location
User
accessible
To clear or
sanitize
CMOS, SDRAM,
128 Mbit x 16, 8 devices,
2Gbea,16G
b total,
DDR2, 1.8 V, 400 MHz
(Effective 800 MHz);
MT47H128
M16RT-25E,
BGA
1
Contains
waveform
main
data
during
operatio
n
Both No Indirect Front panel
control
and oscilloscope
input channels
Main
Acquisition
board
No Remove
power
from the
instrume n
t
for at least
20 seconds
CMOS, SDRAM,
64 Mbit x 16, 8 devices
1 Gb ea, 8
Gb total,
DDR2, 1.8 V, 400 MHz
(Effective 800 MHz);
MT47H6
4M16HR-25E
BGA
2
Contains
waveform
main
data
during
operat
ion
Both No Indirect Front pan
el control
and oscilloscope
input channels
Main
Acquisition
board
No Remove
power
from the
instrum
ent
for at least
20 seconds
CMOS, SDRAM,
64M x 16, 1024 Mb
DDR2,
1.8 V, 400 MHz
(Effective 800 MHz);
MT47H64M16HR-25E
3
Contains
instrument
code
and
data
duri
ng
operation
Both No Indire
ct
Front p
anel control
and oscilloscope
input channels.
Data i
nputs through
USB port and
Ethernet port.
Main
acquisition
board
No Remove
power
from the
instr
ument
for at least
20 seconds
CMOS
, SDRAM,
64M x 16, 1024 Mb
DDR2, 1.8 V, 400 MHz
(Ef
fective 800 MHz);
MT47H64M16HR-25E
4
Cont
ains
display
images
dur
ing
operation
Both No Indirect Front panel control
and o
scilloscope
input channels.
Main
acqu
isition
board
No Remove
powe
r
from the
instrument
for
at least
20 seconds
CMOS, DRAM;
4M X 32,128 Mb,
SD
RAM, 3.3 V;
MT48LC4M32B2B\5-7,
90-BALL FBGA
(
11 MM X 13 MM),
DS4
Contains
code for
di
gital
waveform
processor
No
ne
No In
direct
System
Ma
in
acquisition
board
No Re
move
power
from the
in
strument
for at least
20 seconds
2Declassication and Security Instructions

Other manuals for Tektronix MDO4000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MDO4000B Series and is the answer not in the manual?

Tektronix MDO4000B Series Specifications

General IconGeneral
BrandTektronix
ModelMDO4000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals