Built-In Security Features
To Overwrite Protected
User Data
This procedure
overwrites everything that is stored in the protected user data area.
Send the following comm ands to the oscilloscope:
:PASSWORD “XYZZY” (or current password if changed from the default of
“XYZZY”)
:NEWPASS “ANYTHING”
:PASSWORD “ANYTHING”
*PUD #3 300 xxx xxx xxx xxx xxx xxx xxx xxxx xxx xxx xxxxxxxxxx[repe at
for a total of 300 x’s]
NOTE. If you do not have access to a program that supports sending programmatic
commands to the instrument, copy the preceding commands to a text file that ends
in “.set
” and recall the file from the Recall Setup menu.
For more information on using programming commands, refer to the Programmer
Manual
for this instrument, available at www.tektronix.com/downloads.
10 Declassification and Security Instructions