Built-In Security Features
To Overwrite Protected
User Data
This procedure
overwrites everything that is stored in the protected user data area.
Send the following comm ands to the oscilloscope:
:PASSWORD “XYZZY” (or current password if changed from the default of
“XYZZY”)
:NEWPASS “ANYTHING”
:PASSWORD “ANYTHING”
*PUD #3 300 xxx xxx xxx xxx xxx xxx xxx xxxx xxx xxx xxxxxxxxxx[repe at
for a total of 300 x’s]
NOTE. If you do not have access to a program that supports sending programmatic
commands to the instrument, copy the preceding commands to a text le that ends
in “.set
and recall the le from the Recall Setup menu.
For more information on using programming commands, refer to the Programmer
Manual
for this instrument, available at www.tektronix.com/downloads.