Analyze Waveform Data
Selecting Automatic Measurements
The following tables list each automatic measurement by category: time or amplitude. (See p age 190, Ta ki ng A ut om at ic
Measurements.)
Time Measurements
Measurement Description
Period
The time required to complete the rst cycle in a waveform or gated region. Period is the
reciprocal of frequency and is measured in seconds.
Frequency
The rst cycle in a waveform or gated region. Frequency is the reciprocal of the period; it is
measured in hertz (Hz) where one Hz is one cycle per second.
Delay
The time between the mid reference (default 50%) amplitude point of two different waveform s.
See also Phase.
Rise Time
The time requ ired for the leading edge of the rst pulse in the w aveform or gate d region
to rise from the low reference value (default = 10%) to the high reference value (default
= 90%) of the nal value.
Fall Time
The time required for the falling edge of the rst pulse in the waveform or gated region to fall
from the high reference value (default = 90%) to the low reference value (default = 10%) of
the nal value.
Positive
Duty
Cycle
The ratio of the positive pulse width to the signal period expressed as a percentage . T he duty
cycle is measured on the rst cycle in the waveform or gated region.
192 MSO4000 and DPO4000 Series Oscilloscopes User Manual

Table of Contents

Other manuals for Tektronix MSO4104

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO4104 and is the answer not in the manual?

Tektronix MSO4104 Specifications

General IconGeneral
BrandTektronix
ModelMSO4104
CategoryTest Equipment
LanguageEnglish

Related product manuals