Installation
6. Push OK Compensate Sig nal Paths from
the resulting side-bezel menu.
OK Com-
pensate
Signal
Paths
The calibration will take approximately
10 minutes to complete.
7. After calibration, verify that the status indicator
on the lower-bezel menu displays Pass.
Utility Page
Calibration
Signal Path
Pass
Factory
Pass
If it does not, then recalibrate t he ins trument
or have the instrument serviced by qualified
service personnel.
8. Service personnel use the factory calibration
functions to calibrat e the internal voltage
references of the oscilloscope u sing external
sources. Refer to your Tektronix field office
or representative for assistance with facto ry
calibration.
NOTE. Signal Path Compensation does not include calibration to the probe tip. (See page 23, Compensating a Passive Voltage
Probe.)
MSO4000 and DPO4000 Series Oscilloscopes User Manual 35