EasyManuals Logo
Home>Tektronix>Test Equipment>MSO4104B

Tektronix MSO4104B Technical Reference

Tektronix MSO4104B
87 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #70 background imageLoading...
Page #70 background image
Performance Verication
13. Repeat the proc
edure for all remaining channels as follows:
a. Push the front-panel button to deselect the c hannel that you have already
tested.
b. Move the DC voltage source connection to the next channel input to be
tested.
c. Starting from step 9, repeat the procedure for each channel.
NOTE. 1M t
esting is required for all models.
14. For all models, repeat tests at 1 M impedance as follows:
a. Set the calibrator to 1 M output.
b. Push the front-panel channel 1 button.
c. Set the Termination to 1M.
d. Repeat steps 9 through 13.
62 MSO4000B and DPO4000B Series Specications and Performance Verication

Table of Contents

Other manuals for Tektronix MSO4104B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO4104B and is the answer not in the manual?

Tektronix MSO4104B Specifications

General IconGeneral
BrandTektronix
ModelMSO4104B
CategoryTest Equipment
LanguageEnglish

Related product manuals