Acquisition
Using TekLink and MultiScope Trigger
To get more than 4 channels associated with a single trigger event, get more than 4 channels looking for a problem on a
channel, or build a big AND gate with more than 4 inputs, use TekLink.
NOTE. When using MultiScope Trigger, the B Triggers are not available. TekLink and MultiScope Trigger are not available
on DPO7000, MSO5000, and DPO5000 Series instruments.
Link 2 Instruments
1. Connect two instruments using a TekLink
cable.
2. TekLink connection status is displayed in
the horizontal readout.
3. Select Trig > A Event (Main) Trigger
Setup....
MSO70000/C, DPO/DSA70000B/C, DPO7000, and MSO/DP O5000 Series U ser M anual 63