Install Your Ins
trument
2. Width:
17.96 inches (
456 mm)
17.75 inches (
451 mm)
3. Height:
10.9 inches (2
77 mm)
11.48 inches (
292 mm)
4. Before opera
ting the instrument, verify
the ambient temperature:
5°Cto+45°C(
+41°Fto+113°F)
5°Cto+45°C(
+41°Fto+113°F)
5. Verify the operating humidity: 8% to 80% relative humidity with
a maximum wet-bulb temperature
of +29 °C (+8
4 °F) at or below
+45 °C (+113 °F), noncondensing
Upper limit derated to 30% relative
humidity a
t +45 °C (+113 °F)
8% to 80% relative humidity at up
to+32°C(+90°F)
5% to 45% rel
ative humidity above
+32 °C (+90 °F) up to +45 °C
(+113 °F), noncondensing, and
is limited
by a maximum wet-bulb
temperature of +29.4 °C (+85 °F)
(derates relative humidity to 32%
at +45 °C (
+113 °F) )
6. Verify th
e operating altitude:
<4 GHz mod
els: 3,000 m
(9,843 feet)
≥4GHzmod
els: 3,000 m
(9,843 feet), derate maximum
operating temperature by 1 °C
per 300 m
eters (984.25 feet)
above 1500 meters (4921.25 feet)
altitude.
7. Maximum input voltage, <4 GHz models:
50 Ω
1MΩ
5V
RMS
, with peaks ≤±24 V.
150 V, derate at 20 dB/decade to 9 V
RMS
above 200 kHz. The m aximum
input v
oltage at the BNC, between center conductor and ground is
400 V peak. The RMS voltage is limited to <150 V for arbitrary waveshapes
including DC. The maximum pulse width for impulses with peaks over
150Vi
s50μs. Example: At 0 V to 400 V peak, rectangular wave, the duty
factor is 14%. The maximum transient withstand voltage is ±800 V peak.
MSO70000/C, DPO/DSA70000B/C, DPO7000, and MSO/DP O5000 Series U ser M anual 3