Install Your Ins
trument
Maximum input
voltage, ≥4 GHz models:
50 Ω <1 V
RMS
for <1 V
/FS settings and < 5.5 V
RMS
for ≥1 V/FS settings.
Maximum nondestructive input voltage to
logic probes
, MSO70000 Series:
±15 V
CAUTION. For proper cooling, keep the bottom and sides of the instrument clear of obstructions.
MSO5000 and DPO5000
1. Place the instrument on a cart or bench.
The instrument should rest on its bottom
or rear feet. An optional rack mounting
kit is available. Observe the following
clearance requirements and dimensions:
Rear:
2in(50.8mm)
Left side: 2 in (50.8 mm)
2. Width:
17.3 inches (4
39 mm)
3. Height:
9.2 inches (2
33 mm) including feet
10.7 inches (272 mm) including vertical handle and feet
4. Before operating the instrument, verify
the ambient temperature:
5°Cto+50°C(+41°Fto+131°F)
5. Verify the operating humidity: High: 40 °C to 50 °C (104 °F to 122 °F), 10% to 60% relative humidity
Low: 0 °C to 4
0 °C (32 °F to 104 °F), 10% to 90% relative humidity
6. Verify the
operating altitude:
3,000 m (9,
843 feet)
7. Maximum input voltage:
50 Ω input impedance
1MΩ input
impedance
5V
RMS
, with peaks ≤±20 V (DF ≤6.25%).
300 V
RMS
w
ith peaks ≤±425 V at the BNC.
For <100 mV/div derate at 20 dB/decade above 100 kHz to 30 V
RMS
at
1 MHz, 10 dB/decade above 1 M Hz.
For ≥100
mV/div derate at 20 dB/decade above 3 MHz to 30 V
RMS
at
30 MHz, 10 dB/decade above 30 MHz.
Maximum
nondestructive input voltage to
logic probes:
±42 V
CAUTIO
N. For proper cooling, keep the back and left side of the instrument clear of obstructions.
4 MSO70000/C, DPO/DSA70000B /C, DPO7000, and MSO/DPO5000 Series User Manual