EasyManuals Logo
Home>Tektronix>Test Equipment>P6015A

Tektronix P6015A User Manual

Tektronix P6015A
76 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #29 background imageLoading...
Page #29 background image
In Detail
P6015A Instruction Manual
113
Figure 14: Maximum Input Voltage Derating (DC + Peak AC)
20 kV
28 kV
33 kV
36 kV
40 kV
10% Duty Cycle 100 ms on time
1 sec100 ms
10 ms1 ms
Duration
Maximum Voltage (peak)
20% Duty Cycle 50 ms on time
30% Duty Cycle 40 ms on time
100% Duty Cycle 100 ms on time
50% Duty Cycle 30 ms on time
Figure 15: Peak Pulse Derating
Probe Grounding
When making any measurement, use the probe ground connector to
form the basic two-terminal connection to the device under test.
CAUTION. The ground lead is rated at 1 kV maximum. Make sure the
ground lead does not contact the probe tip or a high-voltage point on
the circuit under test.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix P6015A and is the answer not in the manual?

Tektronix P6015A Specifications

General IconGeneral
BrandTektronix
ModelP6015A
CategoryTest Equipment
LanguageEnglish

Related product manuals