P6015A Instruction Manual
123
Specifications
Warranted Characteristics
This section lists the various warranted characteristics that describe
the P6015A High Voltage Probe. Included are warranted electrical
and environmental characteristics.
Warranted characteristics are described in terms of quantifiable
performance limits which are warranted.
The electrical characteristics listed in Table 13 apply under the
following conditions:
H The probe and instrument with which it is used must have been
calibrated at an ambient temperature of between +20 °C and
+30 °C.
H The probe and instrument must be in an environment whose
limits are described in Table 1-3.
H The probe and instrument must have had a warm-up period of at
least 20 minutes before applying elevated voltages.