The results show the three difference traces displayed in the screen at the same time.
■
The Average trace is generated by multiple traces averaged together. There is one vertical value for each
underlying frequency data point. Once the specified number of traces have been acquired and averaged to
generate the displayed trace, each new trace takes the place of the oldest trace in the calculation.
■
Max Hold trace displays the maximum value in the trace record for each display point. Each new trace display
point is compared to the previous maximum value and the greater value is retained for display and subsequent
comparisons.
■
Min Hold trace displays the minimum value in the trace record for each display point. Each new trace display
point is compared to the previous minimum value and the lesser value is retained for display and subsequent
comparisons.
Demo instructions
34 RSA306 Self-Guided Demo