EasyManuals Logo
Home>Tektronix>Measuring Instruments>RSA6114A

Tektronix RSA6114A User Manual

Tektronix RSA6114A
84 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #51 background imageLoading...
Page #51 background image
Operation
Connecting a Signal
The analyzer has a single R F signal input located on the front panel. M ake sure that you observe the input signal
requirements before connecting a signal to the input.
CAUTION. To prevent static damage to the instrument or to any instrument accessories, discharge to ground any
electrostatic charge that may be present on the center and outer conductors of cables, before attaching the cables to the
instrument inputs. Do not create an ESD (electrical static damage) antenna by leaving cables attached to the instrument with
the other ends of the cables open.
Required Signal Levels
The following table shows the input signal levels over which the analyzer can be used for m easurements. The accuracy is
guaranteed at a signal amplitude and frequency (normalization point). Measurements c an be made on signals within the
entire ran
ge of allowable inputs, but the accuracy of the measurem ent is affected by the frequency and amplitude of the
signal to be measured (see the RSA6100A Series or RSA5100A Series data sheet for more information).
RF signal
input range (CW or peak envelope power)
Modulation measurement requirement S pectral display requirement
-50 dBm t
o +30 dBm
DANL to +
30 dBm
Input si
gnal requirements
Characteristic Description
Input i
mpedance
50
Input frequency range
RSA6106A
RSA61
14A
RSA6120A
RSA5103A
RSA5
106A
RSA5103A/RSA5106A
9kHzto6.2GHz
9kHzto14GHz
9kHz
to 20 GHz
9 kHz to 3 GHz (RF Band)
9 kHz to 6.2 GHz (RF Band)
1Hzt
o 32 MHz (LF Band)
Max
imum measurable continuous input power
1 W (+30 dBm)
Max
imum voltage rating
RSA6100A: ±40 V DC, AC coupled
RSA5100A ±5 V DC, AC coupled
CAUTION. To prevent damage to the instrument, do not connect CW signals greater than 1 Watt (+30 dBm). For the
RSA6100A Series, the maximum voltage rating is ±40 V DC. The maximum pulse specications are 75 W peak, 10 μs pulse
wi
dth, 0.001 duty cycle, and 30 dB attenuation. For the RSA5100A Series, the maximum voltage rating is ±5 V DC. The
maximum pulse specications are 50 W peak, <10 µs pulse width, 0.001 duty cycle and 30 dB attenuation.
RSA6100A Series & RSA5100A Series Q uick Start User Manual 35

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix RSA6114A and is the answer not in the manual?

Tektronix RSA6114A Specifications

General IconGeneral
BrandTektronix
ModelRSA6114A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals