DTM500 Series Adjustment Procedures
Handheld and Benchtop Instruments Basic Service
13
DTM510 Adjustment Procedure
This section describes how to adjust the DTM510 thermometer. To properly
adjust the thermometer, perform the following steps in sequential order.
1. Open the thermometer as described in the previous section.
2. Turn on the thermometer and allow it to warm up for at least 30 minutes.
3. Connect the thermocouple simulator to the probe connector.
4. Set the simulator probe type to K.
5. For each row in Table 7, set the thermometer and simulator as shown; then
adjust the specified test point until the displayed temperature is within the
proper limits. Figure 2 shows the location of the adjustment points.
Table 7: DTM510 Adjustment Procedures
Thermometer Setting Simulator Setting
C/F Resolution Temperature Test Point Temperature Limits
C 0.1_ 0_ C VR1 –00.1_ and 00.1_
F 1_ 1000_ C VR3 1831_ and 1833_
C 1_ 1000_ C VR5 999_ and 1001_
C 0.1_ 190_ C VR2 189.9_ and 190.1_
F 0.1_ 0_ C VR4 31.9_ and 32.1_
C 1_ 1000_ C VR3 999_ and 1001_

Table of Contents

Other manuals for Tektronix TAS 200 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TAS 200 Series and is the answer not in the manual?

Tektronix TAS 200 Series Specifications

General IconGeneral
BrandTektronix
ModelTAS 200 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals