TAS 200 Series Performance Verification
Handheld and Benchtop Instruments Basic Service
7
Test Equipment
The performance verification procedures use external traceable test equipment to
directly check warranted characteristics. If you substitute equipment, always
choose instruments that meet or exceed the minimum requirements specified in
Table 11.
Alternative test equipment must meet or exceed the intended minimum
requirements. If you substitute equipment, you may need to modify the
performance verification procedures.
NOTE. Before beginning the performance verification procedures, warm up the
test equipment according to the manufacturer’s recommendations.
Table 11: Performance Verification Test Equipment
Description Minimum Requirements Example Product
Leveled Sine Wave Generator 200 kHz to 250 MHz; variable amplitude
from 5 mV to 4 V
p-p
into 50 W
Wavetek
1
Universal
alibration
System with Oscilloscope Calibration
Time Mark Generator Variable marker frequency from 10 ms to
10 ns; accuracy within 2 ppm
Module (Option 2
)
Fluke 5500A Multi-product Calibrator with
DC Calibration Generator DC voltage levels from 100 mV to 10 V
Oscilloscope Calibration Option 5500A-SC
Digital Multimeter 35 V, 4 A, ±0.1% Accuracy Tektronix DMM252 Digital Multimeter
Termination (two required) 50 W, female BNC input, male BNC output Tektronix 011-0049-01
Cable, Coaxial 75 W, male-to-male BNC connectors,
36 inch length
Tektronix 012-1338-00
Cable, Precision Coaxial 50 W, male-to-male BNC connectors,
36 inch length
Tektronix 012-0482-00