Page #206 background image
PS280 and PS283 Performance Verification
12
Handheld and Benchtop Instruments Basic Service
Table 9: Load Resistor Values for Voltage Checks
de
PS280 PS283
Mo
de
R
L
R
L
INDEP 15 , 70 W 30 , 40 W
SERIES 30 , 140 W 60 , 70 W
PARALLEL 7.5 , 140 W 15 , 70 W
5V FIXED 1.7 , 20 W 1.7 , 20 W
The following check verifies the SLAVE voltage regulation performance.
1. Set the power supply to the INDEP operating mode by disengaging both
TRACKING mode switches (both switches out).
2. Set up the digital multimeter to measure 30 VDC.
3. Connect the digital multimeter and load resistors to the + and – terminals of
the SLAVE output as shown in Figure 4. See Table 9 for the appropriate load
resistor values.
4. Verify that you can adjust the power supply from 0 to 30 V.
The following check verifies the 5 V fixed voltage regulation performance.
1. Set up the digital multimeter to measure 5 VDC.
2. Connect the digital multimeter and load resistors to the 5V FIXED 3A
terminals. Use the test setup illustrated in Figure 4. See Table 9 for the
appropriate load resistor values.
3. Verify that the power supply maintains an output of 5 V,
±0.25 V
SLAVE Voltage Regulation
5 V Fixed Voltage
Regulation

Table of Contents

Other manuals for Tektronix TAS 200 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TAS 200 Series and is the answer not in the manual?

Tektronix TAS 200 Series Specifications

General IconGeneral
BrandTektronix
ModelTAS 200 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals