TAS 200 Series Performance Verification
8
Bench Test Instruments and Handheld Oscilloscopes Basic Service
Test Equipment
The performance verification procedures use external traceable test equipment to
directly check warranted characteristics.
Alternative test equipment must meet or exceed the intended minimum
requirements. If you substitute equipment, you may need to modify the
performance verification procedures.
NOTE. Before beginning the performance verification procedures, warm up the
test equipment according to the manufacturer’s recommendations.
Table 11: Performance verification test equipment
Description Minimum requirements Example product
Leveled Sine Wave Generator 200 kHz to 250 MHz; variable amplitude
from5mVto4V
p-p
into 50 Ω
Wavetek 9100 Universal Calibration
System with Oscilloscope Calibration
Time Mark Generator Variable marker frequency from 10 ms to
10 ns; accuracy within 2 ppm
Module (Opti on 250)
Fluke 5500
Multi-produ
tC
libr
tor with
DC Calibration Generator DC voltage level s f rom 100 mV to 10 V
Oscilloscope Calibration Option 5500A-SC
Digital Multimeter 35 V, 4 A, ±0.1% Accuracy Tektronix DMM914 Digit al Mul timeter
Termination (two required) 50 Ω, female BNC i nput, male BNC output Tektronix 011-0049-01
Cable, Coaxial 75 Ω, male-to-male BNC connect ors,
36 inch length
Tektronix 012-1338-00
Cable, Precisi on Coaxi al 50 Ω, male-to-male BNC connect ors,
36 inch length
Tektronix 012-0482-00